Team:Exeter/EnzymeValidation

From 2014.igem.org

(Difference between revisions)
 
(37 intermediate revisions not shown)
Line 10: Line 10:
</script>
</script>
</head>
</head>
-
<h1>Validation of Enzyme Function and Activity</h1>
 
-
<h2>Summary</h2>
 
-
<p>This experiment shows that Raman spectroscopy can be used to measure the concentration of NG in aqueous solutions. It was shown to provide accurate quantifiable measurements of the concentration of NG where the main source of error was in the preparation of the samples themselves.</p>
 
-
<p>However the use of Raman spectroscopy to measure low concentrations of artificial compounds in the presence of organic material is not recommended. The issue with measuring concentrations in the presence of cells is the large amount of different systems inside the cell, these produced a large fingerprint region (800cm$^{-1}$ to 1800cm$^{-1}$) that can mask the peaks of the artificial compound. The cells had a large auto-florescence which contributed to the difficulty of extracting useful information from the spectrum.</p>
+
<table id="toc" class="toc">
 +
<tr><td>
 +
<div id="toctitle"><h2>Contents</h2></div>
 +
<ul>
 +
<li class="toclevel-1"><a href="#1"><span class="tocnumber">1.</span> <span class="toctext">Summary</span></a></li>
-
<p>Raman spectroscopy is better suited to measuring concentrations of nitroglycerin and TNT <i>in vitro</i> where higher concentrations can be used and auto-florescence can be kept to a minimum. For measurements <i>in vivo</i> the accuracy and rate of gathering data could be greatly improved by adding an nonreactive marker compound. This compound would have a known peak area and concentration and would allow for the calibration of other peaks present in a spectrum. This method could remove uncertainties in focusing the laser.</p>
+
<li class="toclevel-1"><a href="#2"><span class="tocnumber">2.</span> <span class="toctext">Abstract</span></a></li>
-
<p>
+
<li class="toclevel-1"><a href="#3"><span class="tocnumber">3.</span> <span class="toctext">Results</span></a></li>
-
NemA is capable of catalysing the conversion of Nitroglycerin to various products using NADH and FMN as cofactors. The binding affinity of the protein for this substrate (the Michealis Menten constant, Km) and the maximum reaction velocity (Vmax), were determined and are comparable to the published values; shown in figure 1. NemA is therefore a suitable enzyme for use in our system and has been shown to function at the physiologically relevant pH of 7.  
+
<ul>
-
</p>
+
<li class="toclevel-2"><a href="#3.1"><span class="tocnumber">3.1</span> <span class="toctext">The Raman spectrum of TNT and nitroglycerin</span></a></li>
 +
<li class="toclevel-2"><a href="#3.2"><span class="tocnumber">3.2</span> <span class="toctext">The Standard curve</span></a></li>
 +
<li class="toclevel-2"><a href="#3.3"><span class="tocnumber">3.3</span> <span class="toctext">NemA Activity</span></a></li>
 +
<li class="toclevel-2"><a href="#3.4"><span class="tocnumber">3.4</span> <span class="toctext">NemA Kinetic Analysis</span></a></li>
 +
</ul>
-
<h2>Abstract</h2>  
+
<li class="toclevel-1"><a href="#4"><span class="tocnumber">4.</span> <span class="toctext">Eperimental Method</span></a></li>
-
<p>XenB is a protein that the Exeter iGEM team propose will allow E.coli to degrade TNT and Nitroglycerin, at concentrations above those normally toxic to the cell. Among many others proposed, NemA catalyses the reaction shown in figure 2.
+
 +
<ul>
 +
<li class="toclevel-2"><a href="#4.1"><span class="tocnumber">4.1</span> <span class="toctext">The Standard Curve</span></a></li>
 +
<li class="toclevel-2"><a href="#4.2"><span class="tocnumber">4.2</span> <span class="toctext">Purifying NemA</span></a></li>
 +
<li class="toclevel-2"><a href="#4.3"><span class="tocnumber">4.3</span> <span class="toctext">Stopped Enzyme Assay</span></a></li>
 +
</ul>
 +
<li class="toclevel-1"><a href="#5"><span class="tocnumber">5.</span> <span class="toctext">References</span></a></li>
 +
</ul>
 +
</td></tr>
 +
</table>
 +
<script>if (window.showTocToggle) { var tocShowText = "show"; var tocHideText = "hide"; showTocToggle(); } </script>
-
<p>The aim of this experiment was to measure the degradation rate of trinitrotoluene (TNT) and nitroglycerin (NG) <i>in vitro</i> by pure samples of XenB.</p>
+
<h1><span id="0">Validation of Enzyme Activity</span></h1>
 +
<h2><span id="1">Summary</span></h2>
-
<b>Raman Spectroscopy </b>
+
<p>In these experiments we demonstrate that the NemA protein (encoded by <a href="http://parts.igem.org/Part:BBa_K1398002">BBa_K1398002</a>) expressed as part of the construct <a href="http://parts.igem.org/Part:BBa_K1398003">BBa_K1398003</a> can catalyse the degradation of nitroglycerin. Analysis of the data collected allowed the kinetic parameters of Km and Vmax of NemA for nitroglycerin to be calculated as 6 mM and 21 umol mg$^{-1}$ min$^{-1}$ respectively. To demonstrate this we have used Raman spectroscopy to measure the concentration of nitroglycerin in aqueous solutions. For TNT please see the HPLC page.</p>
-
<p>This is a technique used to observe vibrational, rotational and other low frequency modes in a system. It utilizes the inelastic scattering of monochromatic light from these low frequency modes by detecting the shift in reflected photon energy. Inelastic scattering occurs where part of the photons energy is transferred to the vibrational state resulting in a lower energy reflected photon. This shift can be used to give information about the types of bonds in the system and how many of these are present.</p>
+
-
<p>Typically the inelastically reflected light from the low frequency modes in a system is very weak in comparison to the elastically scattered light from the sample. In the past this had been the major barrier to the wider use of Raman spectroscopy. With recent improvements to band-stop filters the ease of processing the signal from the sample has been greatly improved and as a result Raman spectroscopy has been put to greater use.</p>
 
-
<p>Raman spectroscopy was used for this experiment because of its ability to both detect the compounds of interest and give quantifiable values for their concentration. In addition, this experiment will provide an indication as to whether this type of spectroscopy is suitable for use on aqueous organic samples.</p>
+
<h2><span id="2">Abstract</span></h2>
 +
<p>NemA is a protein the Exeter iGEM team propose will allow <i>E. coli</i> to degrade TNT and nitroglycerin at concentrations above those normally toxic to the cell. The following experimental account describes the protocol used to confirm the substrate target of NemA and analyse the kinetic capability of this enzyme.</p>
 +
<p>Raman spectroscopy is a technique used to observe vibrational, rotational and other low frequency modes in a system. It utilises the inelastic scattering of monochromatic light from these low frequency modes by detecting the shift in reflected photon energy. Inelastic scattering occurs where part of the photons energy is transferred to the vibrational state resulting in a lower energy reflected photon. This shift can be used to give information about the types of bonds in the system and how many of these are present.</p>
 +
<p>Typically the inelastically reflected light from the low frequency modes in a system is very weak in comparison to the elastically scattered light from the sample. In the past this had been the major barrier to the wider use of Raman spectroscopy. With recent improvements to band-stop filters the ease of processing the signal from the sample has been greatly improved and as a result Raman spectroscopy has been put to greater use.</p>
 +
<p>Raman spectroscopy was used for this experiment because of its ability to both detect the compounds of interest and give quantifiable values for their concentration. The data obtained in this experiment  complements similar measurements made using high performance liquid chromatography. In addition, this experiment will provide an early indication as to whether this type of spectroscopy is suitable for use on aqueous organic samples.</p>
-
<h2>Results, Analysis and Discussion</h2>
+
<h2><span id="3">Results</span></h2>
 +
<h3><span id="3.1">The Raman spectrum of TNT and nitroglycerin</span></h3>
-
<b>Choice of analytical procedure</b>
+
<p>Figure 1 shows the Raman spectrum of NG and TNT observed using undiluted samples. The NG spectrum agrees with literature spectrum having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$ [3]. The TNT spectrum corresponds to the literature spectrum [2] with its highest intensity peaks at 820$^{-1}$ and 1360 cm$^{-1}$.</p>
-
<p>NADH is a cofactor in the breakdown of TNT and nitroglycerin by NemA. This opens up the possibility of simply measuring the catalytic rate of nitroglycerin or TNT degradation by following the change in absorbance at 340nm. However, very early in the process we realised that both TNT and nitroglycerin also produce a significant absorbance at 340nm; a fact that would complicate our analytical procedure. Raman spectroscopy was hypothesised to be capable of isolating the characteristic signals of nitroglycerin and TNT from a complex mixture, allowing quantification of these compounds.</p>
+
-
 
+
-
<b>The Raman Spectrum of TNT and Nitroglycerin</b>
+
-
 
+
-
<p>Figure 2 shows the Raman spectrum of NG and TNT observed using undiluted samples. The NG spectrum agrees with literature spectrum having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$ [3]. The TNT spectrum corresponds to the literature spectrum [2] with its highest intensity peaks at 820$^{-1}$ and 1360 cm$^{-1}$.</p>
+
-
 
+
-
<p>The TNT spectrum has much lower intensity peaks compared to that of nitroglycerin. This is to be expected as the TNT sample was at a lower concentration (1mgml$^{-1}$) compared to that of Nitroglycerin (2mgml$^{-1}$).</p>
+
 +
<p>The TNT spectrum has a low intensity in compared to the nitroglycerin. This is to be expected as the nitroglycerin sample was at a higher concentration (2mgml$^{-1}$) than the TNT sample (1mgml$^{-1}$).</p>
<figure>
<figure>
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/9/97/Exeter2014_Raman_Spectrum.png">
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/9/97/Exeter2014_Raman_Spectrum.png">
-
<figcaption>Figure 2: The Raman spectrum of  TNT and nitroglycerin. The TNT spectrum agrees with literature values having high intensity peaks at 820cm$^{-1}$ and 1360cm$^{-1}$ [2]. Nitroglycerin also agrees with literature values having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$[3].</figcaption></figure>
+
<figcaption>Figure 1: The Raman spectrum of  TNT and nitroglycerin. The TNT spectrum agrees with literature values having high intensity peaks at 820cm$^{-1}$ and 1360cm$^{-1}$ [2]. Nitroglycerin also agrees with literature values having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$[3].</figcaption>
 +
</figure>
-
<p>Figure 3 shows the normalized Raman spectrum for nitroglycerin at various concentrations between the wavenumbers of 850 cm$^{-1}$ and 750 cm$^{-1}$. The spectrum was normalized between 1200 cm$^{-1}$ and 600 cm$^{-1}$ by dividing by the average intensity between these points. The concentration is proportional to the peak area.</p>
+
<h3><span id="3.2">The Standard curve</span></h3>
 +
<p>Figure 2 shows the normalised Raman spectrum for nitroglycerin at various concentrations between the wavenumbers of 850 cm$^{-1}$ and 750 cm$^{-1}$. The spectrum was normalised between 1200 cm$^{-1}$ and 600 cm$^{-1}$ by dividing by the average intensity between these points. The concentration is proportional to the peak area.</p>
<figure>
<figure>
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/5/56/Exeter2014_Raman_DoublePeak.png">
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/5/56/Exeter2014_Raman_DoublePeak.png">
-
<figcaption>Figure 3: The normalized Raman spectrum for nitroglycerin. The concentration of nitroglycerin is proportional to the peak area.</figcaption>
+
<figcaption>Figure 2: The normalised Raman spectrum for nitroglycerin. The concentration of nitroglycerin is proportional to the peak area.</figcaption>
</figure>
</figure>
-
<p>The figure 3 also shows the predicted background intensity level without the peaks. This was made by plotting a least square fitted line through measurements around the peak sites. The background intensity level increases with lower intensity peaks, this is an artifact of normalization.</p>
+
<p>The figure 2 also shows the predicted background intensity level without the peaks. This was made by plotting a least square fitted line through measurements around the peak sites. The background intensity level increases with lower intensity peaks, this is an artifact of normalisation.</p>
<p>Using the predicted background intensity the peak area, and thereby the concentration, can be calculated using the trapezium rule,</p>
<p>Using the predicted background intensity the peak area, and thereby the concentration, can be calculated using the trapezium rule,</p>
Line 79: Line 93:
<p>Where $I_i$ is the intensity of the $i$th measurement and $k_i$ is the wavenumber of the $i$th measurement.</p>
<p>Where $I_i$ is the intensity of the $i$th measurement and $k_i$ is the wavenumber of the $i$th measurement.</p>
-
<p>The peak area is plotted against the concentration of the prepared sample in figure 4. A least square fitted line (Red dashed line) does not intercept the origin of the graph. This error could be due to a systematic error in transferring the nitroglycerin into the dilutions. The stock nitroglycerin solution was quite viscous meaning that some extra nitroglycerin stuck to the outside of the pipette, this would have increased the amount of compound in each dilution.</p>
+
<p>The peak area is plotted against the concentration of the prepared sample in figure 3. A least square fitted line (Red dashed line) does not intercept the origin of the graph. This error could be due to a systematic error in transferring the nitroglycerin into the dilutions. The stock nitroglycerin solution was quite viscous meaning that some extra nitroglycerin stuck to the outside of the pipette, this would have increased the amount of compound in each dilution.</p>
-
<p>The error in the NG concentration attempted to account for the difficulty of mixing the correct quantities of NG and water by estimating the quantity of nitroglycerin to be $\pm$ 0.5 $\mu$l. This is probably incorrect as the actual error would be biased towards greater concentrations of nitroglycerin.</p>
+
<p>The error in the nitroglycerin concentration attempted to account for the difficulty of mixing the correct quantities of NG and water by estimating the quantity of nitroglycerin to be $\pm$ 0.5 $\mu$l. This is probably incorrect as the actual error would be biased towards greater concentrations of nitroglycerin.</p>
-
<p>Assuming that the peak height must be zero at zero concentration another least square fitted line was plotted, this time fixing the intercept to zero. This line will be used for determining concentrations of nitroglycerin in future measurements.</p>
+
<p>Assuming that the peak height must be zero at zero concentration another least square fitted line was plotted, this time fixing the intercept to zero. This line will be used for determining concentrations of NG in future measurements.</p>
<figure>
<figure>
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/4/4c/Exeter2014_Raman_Standard.png">
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/4/4c/Exeter2014_Raman_Standard.png">
-
<figcaption>Figure 4: Peak area of Raman spectrum for known concentrations of nitroglycerin. The nitroglycerin concentrations are probably lower than the real values due to a systematic error incurred by the viscosity of the stock nitroglycerin solution.</figcaption>
+
<figcaption>Figure 3: Peak area of Raman spectrum for known concentrations of nitroglycerin. The nitroglycerin concentrations are probably lower than the real values due to a systematic error incurred by the viscosity of the stock nitroglycerin solution.</figcaption>
</figure>
</figure>
Line 100: Line 114:
<p>No standard curve for TNT was made. TNT was found to be immiscible with water resulting in extremely variable peak intensities when measuring TNT/water samples. This was likely due to the TNT forming a layer on top of the water, when part this layer was deposited onto the slide the concentration would be unexpectedly high. However for most measurements only the water part of the sample would be placed onto the slide resulting in no peaks.</p>
<p>No standard curve for TNT was made. TNT was found to be immiscible with water resulting in extremely variable peak intensities when measuring TNT/water samples. This was likely due to the TNT forming a layer on top of the water, when part this layer was deposited onto the slide the concentration would be unexpectedly high. However for most measurements only the water part of the sample would be placed onto the slide resulting in no peaks.</p>
-
<p>Due to the immiscibility of TNT and water, measurements of TNT concentration could not be made using Raman spectroscopy. Please see our high performance liquid chromatography page for a proposed solution to this problem.</p>
+
<p>Due to the immiscibility of TNT and water, measurements of TNT's concentration could not be made.</p>
-
<b>Determination of NemA activity</b>
+
<h3><span id="3.3">NemA Activity</span></h3>
-
<p>Figure 5 shows NemA reaction kinetics as a function of initial velocity, Vi, against initial TNT concentration (derived from taking a tangent at the steepest section of each series in figure 4). From this it can be concluded that the Vmax is...</p>
+
<p>The results of the control experiments (lacking a single cofactor or NemA) are shown in figure X.  
<figure>
<figure>
-
<img class="https://static.igem.org/mediawiki/2014/1/15/NemA_initial_velocities.png</img>
+
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/e/ed/NG_Controls.jpg"/>
-
<figcaption>Figure 5: NemA reaction kinetics as a function of initial velocity, Vi, against initial TNT concentration<figcaption>
+
<figcaption>Figure 4: Nitroglycerin concentration does not change unless FMN, NADH, NADPH and NemA are present.</figcaption>
</figure>
</figure>
-
<p>
+
<p>These results show that all three cofactors are required for NemA to catalyse the degradation of nitroglycerin. Furthermore NemA is responsible for the decrease in nitroglycerin concentration shown in figure X.
-
The data can either be plotted as a Lineweaver-Burke plot (figure 6), or as a Hanes plot (figure 7) to more accurately determine the Km values. From these it can be concluded that the Km of NemA for TNT substrate is...
+
-
</p>
+
 +
<figure>
 +
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/6/67/NemA_degradation_of_nitroglycerin_kinetics_exp.png"/>
 +
<figcaption>Figure 5: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin. </figcaption>
 +
</figure>
 +
<p>Figure 5 shows that iGEM Exeter have produced a synthetic <i>E. coli</i> bacterial strain which produces a protein capable of degrading nitroglycerin!</p>
 +
<h3><span id="3.4">NemA Kinetic Analysis</span></h3>
 +
<p>Figure 6 shows NemA reaction kinetics as a function of initial velocity, Vi, against initial TNT concentration (derived from taking a tangent at the steepest section of each series in figure 5).</p>
 +
<figure>
 +
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/e/e9/NemA_degradation_of_nitroglycerin_kinetics_exp_Vi_vs_s.png"/>
 +
<figcaption>Figure 6: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin. </figcaption>
 +
</figure>
 +
<p>The data can be displayed as a Hanes plot (figure 7) to accurately determine the Km and Vmax values.</p>
 +
<figure>
 +
<img class="large_centre_image" src="https://static.igem.org/mediawiki/2014/f/f8/Hanes.png"/>
 +
<figcaption>Figure 6: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin. </figcaption>
 +
</figure>
-
<h2>Experimental Method</h2>
+
<p>Analysis of these data allowed the kinetic parameters of Km and Vmax of NemA for nitroglycerin to be calculated:</p>
-
<b>Qualification of TNT and Nitroglycerin</b>
+
<ul>
 +
<li>Km = 6 mM</li>
 +
<li>Vmax = 21 umol mg$^{-1}$ min$^{-1}$</li>
 +
</ul>
-
<p>The spectra of TNT and Nitroglycerin samples were observed by placing 1 μl of the sample solution onto an aluminium slide and the slide placed into the Raman microscope. The microscope acts as a sight for the laser; a cross hair on the microscope mounted camera shows where the laser will become incident on the droplet.</p>
+
<p>(0.046 mg of NemA was present in each reaction mixture)</p>
-
<p>A 785nm laser was chosen to detect the carbon nitrogen bonds found in nitroglycerin and TNT [3]. The laser was set to a power output of 33 Watts, this was a compromise between achieving high peak intensities and preventing large convection currents in the droplet from occuring. Inelastic scattering was scanned between wavenumbers of 2000nm$^{-1}$ and 500nm$^{-1}$. Two scans were completed and combined to accentuate any peaks.</p>
+
<p>These values were used in the mathematical model we have devleoped which can be seen on the <a href="https://2014.igem.org/Team:Exeter/Modelling">Modelling </a>page.</p>
-
<p>This preliminary testing allowed the characteristic peaks of each compound to be observed and compared to published values. For details, see the results section.</p>
 
-
<b>Quantifying the degradation of TNT and Nitroglycerin</b>
 
-
<p>Pure samples of NemA were achieved by....</p>
 
-
<p>Purified NemA protein was assayed for its ability to degrade Nitroglycerin at a range of concentrations: from 0 mM up to X mM. The reaction is a stopped enzyme assay. In this system the reaction is started by the addition of Nitroglycerin to a reaction mix containing _mM NemA, _mM FMN and _mM NADH. The reaction is stopped by boiling samples at specific time points and the concentration of Nitroglycerin is then assayed using the Raman Spectroscopy method described in the "Qualification of TNT and Nitroglycerin" section.<p>
 
 +
<h2><span id="4">Experimental Method</span></h2>
 +
<h3><span id="4.1">The Standard Curve</span></h3>
 +
<p>In order for the results of any concentration measurement to be analyzed a standard curve of compound concentration versus peak area must be created. To do this the compound is mixed from a known concentration with distilled water to create solutions at a variety concentrations. For nitroglycerin the concentrations were 200, 100, 66.7, 40 and 20 $\mu$gml$^{-1}$.</p>
 +
<p>To measure the spectra of the compound at different concentrations 1 μl of the solution was placed onto an aluminium slide. The slide was placed into the Raman microscope. The microscope acts as the sight for the laser, a cross hair on the microscope mounted camera output shows where the laser will become incident on the droplet.</p>
 +
 +
<p>A 785nm laser was chosen to detect the carbon nitrogen bonds found in nitroglycerin and TNT [3]. The laser was set to a power output of 33 Watts, this was a compromise between peak intensity and preventing large convection currents in the droplet from building up. Inelastic scattering was scanned between wavenumbers of 2000nm$^{-1}$ and 500nm$^{-1}$. Two scans were completed and combined to accentuate any peaks.</p>
 +
 +
<h3><span id="0">Purifying NemA</span></h3>
 +
<p>A 15 mL overnight culture of Exeter iGEM Team's construct <a href="http://parts.igem.org/Part:BBa_K1398003">BBa_K1398003</a>  was grown in LB over 12 hours from glycerol stock. A 10 mL sample of this overnight culture was added to 500 mL LB and incubated for 3 hours at 37 $^{\circ}$C and 200 RPM, whereupon the optical density of the culture reached 0.7. 0.5 mL IPTG was added. After four hours the culture was centrifuged at 4500 g for 20 minutes at 4 $^{\circ}$C. The resulting pellet was resuspended in Tris buffer and lysed by sonication. The whole cell lysate was centrifuged for a further 30 minutes at 4500 g to remove insoluble components. NemA was separated from the supernatent by passing the whole cell lysate through a Nickel column. NemA as encoded by part <a href="http://parts.igem.org/Part:BBa_K1398003">BBa_K1398003</a> bears a His-tag to enable Ni purification. Purified NemA was stored at 5 $^{\circ}$C and the concentration of the resulting aqueous protein was quantified by absorbance in a Qubit 2.0 Flurometer to be 0.045 mM. </p>
 +
 +
<h3><span id="4.2">Stopped Enzyme Assay</span></h3>
 +
<p>A master mix of FMN (0.2 mM), NADH (0.2 mM), NADPH (0.2 mM) and NemA (0.009 mM) was created. Each reaction was started by addition of nitroglycerin, incubated at 20 $^{\circ}$C, to 0.131 mL of master mix. Final nitroglycerin reaction concentrations were: 9.5 mM, 7.2 mM, 4.8 mM and 2.4 mM. Samples of these reaction mixtures were taken and enzymes denatured by boiling at 1, 5, 10 and 30 minutes after addition of nitroglycerin. In conjunction with these reactions, four control experiments were also performed; involving the single removal of either FMN, NADH, NADPH and NemA. These controls demonstrated nitroglycerin degradation occurred via NemA activity.</p>
 +
 +
<p>Nitroglycerin samples were immediately sampled by Raman spectroscopy; using the method previously described.</p>
 +
 +
 +
 +
 +
 +
 +
<h2><span id="5">References</span></h2>
 +
 +
<ol type="1">
 +
 +
<li>COLLANDER, R., AND BARLUND, H.: Acta Botan. Fennica 11, 1 (1930)</li>
 +
<li>Trace level detection and identification of nitro-based explosives by surface-enhanced Raman spectroscopy, S. Botti,* S. Almaviva</li>
 +
<li>Explosive Detection using the D3 Klarite SERS technology, D3 Technologies Ltd.</li>
 +
<li>Fluorescence Background Problem in Raman Spectroscopy, Shan Yang and Ozan Akkus</li>
 +
 +
</ol>
<h2> Navigation </h2>
<h2> Navigation </h2>
 +
<p><a href="https://2014.igem.org/Team:Exeter/DegradationConstructs">Previous: The Enzymes </a></p>
<p><a href="https://2014.igem.org/Team:Exeter/DegradationConstructs">Previous: The Enzymes </a></p>
-
<p><a href="https://2014.igem.org/Team:Exeter/invivoactivity">Next: <i>in vivo</i> activity </a></p>
+
<p><a href="https://2014.igem.org/Team:Exeter/invivoactivity">Next: <i>in vivo</i>: Raman </a></p>
</div>
</div>
 +
</html>
</html>
{{ExeterFooter}}
{{ExeterFooter}}

Latest revision as of 03:25, 18 October 2014

Exeter | ERASE

Contents

Validation of Enzyme Activity

Summary

In these experiments we demonstrate that the NemA protein (encoded by BBa_K1398002) expressed as part of the construct BBa_K1398003 can catalyse the degradation of nitroglycerin. Analysis of the data collected allowed the kinetic parameters of Km and Vmax of NemA for nitroglycerin to be calculated as 6 mM and 21 umol mg$^{-1}$ min$^{-1}$ respectively. To demonstrate this we have used Raman spectroscopy to measure the concentration of nitroglycerin in aqueous solutions. For TNT please see the HPLC page.

Abstract

NemA is a protein the Exeter iGEM team propose will allow E. coli to degrade TNT and nitroglycerin at concentrations above those normally toxic to the cell. The following experimental account describes the protocol used to confirm the substrate target of NemA and analyse the kinetic capability of this enzyme.

Raman spectroscopy is a technique used to observe vibrational, rotational and other low frequency modes in a system. It utilises the inelastic scattering of monochromatic light from these low frequency modes by detecting the shift in reflected photon energy. Inelastic scattering occurs where part of the photons energy is transferred to the vibrational state resulting in a lower energy reflected photon. This shift can be used to give information about the types of bonds in the system and how many of these are present.

Typically the inelastically reflected light from the low frequency modes in a system is very weak in comparison to the elastically scattered light from the sample. In the past this had been the major barrier to the wider use of Raman spectroscopy. With recent improvements to band-stop filters the ease of processing the signal from the sample has been greatly improved and as a result Raman spectroscopy has been put to greater use.

Raman spectroscopy was used for this experiment because of its ability to both detect the compounds of interest and give quantifiable values for their concentration. The data obtained in this experiment complements similar measurements made using high performance liquid chromatography. In addition, this experiment will provide an early indication as to whether this type of spectroscopy is suitable for use on aqueous organic samples.

Results

The Raman spectrum of TNT and nitroglycerin

Figure 1 shows the Raman spectrum of NG and TNT observed using undiluted samples. The NG spectrum agrees with literature spectrum having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$ [3]. The TNT spectrum corresponds to the literature spectrum [2] with its highest intensity peaks at 820$^{-1}$ and 1360 cm$^{-1}$.

The TNT spectrum has a low intensity in compared to the nitroglycerin. This is to be expected as the nitroglycerin sample was at a higher concentration (2mgml$^{-1}$) than the TNT sample (1mgml$^{-1}$).

Figure 1: The Raman spectrum of TNT and nitroglycerin. The TNT spectrum agrees with literature values having high intensity peaks at 820cm$^{-1}$ and 1360cm$^{-1}$ [2]. Nitroglycerin also agrees with literature values having two high intensity peaks at 800cm$^{-1}$ and 840cm$^{-1}$[3].

The Standard curve

Figure 2 shows the normalised Raman spectrum for nitroglycerin at various concentrations between the wavenumbers of 850 cm$^{-1}$ and 750 cm$^{-1}$. The spectrum was normalised between 1200 cm$^{-1}$ and 600 cm$^{-1}$ by dividing by the average intensity between these points. The concentration is proportional to the peak area.

Figure 2: The normalised Raman spectrum for nitroglycerin. The concentration of nitroglycerin is proportional to the peak area.

The figure 2 also shows the predicted background intensity level without the peaks. This was made by plotting a least square fitted line through measurements around the peak sites. The background intensity level increases with lower intensity peaks, this is an artifact of normalisation.

Using the predicted background intensity the peak area, and thereby the concentration, can be calculated using the trapezium rule,

\begin{equation} \text{Area} = \sum_i^{N-1} \frac{I_i+I_{i+1}}{2}(k_{i+1}-k_i). \end{equation}

Where $I_i$ is the intensity of the $i$th measurement and $k_i$ is the wavenumber of the $i$th measurement.

The peak area is plotted against the concentration of the prepared sample in figure 3. A least square fitted line (Red dashed line) does not intercept the origin of the graph. This error could be due to a systematic error in transferring the nitroglycerin into the dilutions. The stock nitroglycerin solution was quite viscous meaning that some extra nitroglycerin stuck to the outside of the pipette, this would have increased the amount of compound in each dilution.

The error in the nitroglycerin concentration attempted to account for the difficulty of mixing the correct quantities of NG and water by estimating the quantity of nitroglycerin to be $\pm$ 0.5 $\mu$l. This is probably incorrect as the actual error would be biased towards greater concentrations of nitroglycerin.

Assuming that the peak height must be zero at zero concentration another least square fitted line was plotted, this time fixing the intercept to zero. This line will be used for determining concentrations of NG in future measurements.

Figure 3: Peak area of Raman spectrum for known concentrations of nitroglycerin. The nitroglycerin concentrations are probably lower than the real values due to a systematic error incurred by the viscosity of the stock nitroglycerin solution.

The error in the peak area is small, it is calculated via,

\begin{equation} \delta \text{Area} = \sigma \langle k_{i+1}-k_i \rangle \sqrt{\frac{N}{2}}, \end{equation}

where $\sigma$ is the standard deviation from the predicted background intensity of the background intensity measurements. $\langle k_{i+1}-k_i \rangle$ is the average wavenumber gap between measurements and $N$ is the number of measurements. The error in the peak area diverges from the measured area considerably for the large numbers of measurements used to calculate it.

No standard curve for TNT was made. TNT was found to be immiscible with water resulting in extremely variable peak intensities when measuring TNT/water samples. This was likely due to the TNT forming a layer on top of the water, when part this layer was deposited onto the slide the concentration would be unexpectedly high. However for most measurements only the water part of the sample would be placed onto the slide resulting in no peaks.

Due to the immiscibility of TNT and water, measurements of TNT's concentration could not be made.

NemA Activity

The results of the control experiments (lacking a single cofactor or NemA) are shown in figure X.

Figure 4: Nitroglycerin concentration does not change unless FMN, NADH, NADPH and NemA are present.

These results show that all three cofactors are required for NemA to catalyse the degradation of nitroglycerin. Furthermore NemA is responsible for the decrease in nitroglycerin concentration shown in figure X.

Figure 5: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin.

Figure 5 shows that iGEM Exeter have produced a synthetic E. coli bacterial strain which produces a protein capable of degrading nitroglycerin!

NemA Kinetic Analysis

Figure 6 shows NemA reaction kinetics as a function of initial velocity, Vi, against initial TNT concentration (derived from taking a tangent at the steepest section of each series in figure 5).

Figure 6: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin.

The data can be displayed as a Hanes plot (figure 7) to accurately determine the Km and Vmax values.

Figure 6: Decrease in nitroglycerin concentration over time at different starting concentrations of nitroglycerin.

Analysis of these data allowed the kinetic parameters of Km and Vmax of NemA for nitroglycerin to be calculated:

  • Km = 6 mM
  • Vmax = 21 umol mg$^{-1}$ min$^{-1}$

(0.046 mg of NemA was present in each reaction mixture)

These values were used in the mathematical model we have devleoped which can be seen on the Modelling page.

Experimental Method

The Standard Curve

In order for the results of any concentration measurement to be analyzed a standard curve of compound concentration versus peak area must be created. To do this the compound is mixed from a known concentration with distilled water to create solutions at a variety concentrations. For nitroglycerin the concentrations were 200, 100, 66.7, 40 and 20 $\mu$gml$^{-1}$.

To measure the spectra of the compound at different concentrations 1 μl of the solution was placed onto an aluminium slide. The slide was placed into the Raman microscope. The microscope acts as the sight for the laser, a cross hair on the microscope mounted camera output shows where the laser will become incident on the droplet.

A 785nm laser was chosen to detect the carbon nitrogen bonds found in nitroglycerin and TNT [3]. The laser was set to a power output of 33 Watts, this was a compromise between peak intensity and preventing large convection currents in the droplet from building up. Inelastic scattering was scanned between wavenumbers of 2000nm$^{-1}$ and 500nm$^{-1}$. Two scans were completed and combined to accentuate any peaks.

Purifying NemA

A 15 mL overnight culture of Exeter iGEM Team's construct BBa_K1398003 was grown in LB over 12 hours from glycerol stock. A 10 mL sample of this overnight culture was added to 500 mL LB and incubated for 3 hours at 37 $^{\circ}$C and 200 RPM, whereupon the optical density of the culture reached 0.7. 0.5 mL IPTG was added. After four hours the culture was centrifuged at 4500 g for 20 minutes at 4 $^{\circ}$C. The resulting pellet was resuspended in Tris buffer and lysed by sonication. The whole cell lysate was centrifuged for a further 30 minutes at 4500 g to remove insoluble components. NemA was separated from the supernatent by passing the whole cell lysate through a Nickel column. NemA as encoded by part BBa_K1398003 bears a His-tag to enable Ni purification. Purified NemA was stored at 5 $^{\circ}$C and the concentration of the resulting aqueous protein was quantified by absorbance in a Qubit 2.0 Flurometer to be 0.045 mM.

Stopped Enzyme Assay

A master mix of FMN (0.2 mM), NADH (0.2 mM), NADPH (0.2 mM) and NemA (0.009 mM) was created. Each reaction was started by addition of nitroglycerin, incubated at 20 $^{\circ}$C, to 0.131 mL of master mix. Final nitroglycerin reaction concentrations were: 9.5 mM, 7.2 mM, 4.8 mM and 2.4 mM. Samples of these reaction mixtures were taken and enzymes denatured by boiling at 1, 5, 10 and 30 minutes after addition of nitroglycerin. In conjunction with these reactions, four control experiments were also performed; involving the single removal of either FMN, NADH, NADPH and NemA. These controls demonstrated nitroglycerin degradation occurred via NemA activity.

Nitroglycerin samples were immediately sampled by Raman spectroscopy; using the method previously described.

References

  1. COLLANDER, R., AND BARLUND, H.: Acta Botan. Fennica 11, 1 (1930)
  2. Trace level detection and identification of nitro-based explosives by surface-enhanced Raman spectroscopy, S. Botti,* S. Almaviva
  3. Explosive Detection using the D3 Klarite SERS technology, D3 Technologies Ltd.
  4. Fluorescence Background Problem in Raman Spectroscopy, Shan Yang and Ozan Akkus

Navigation

Previous: The Enzymes

Next: in vivo: Raman

Exeter | ERASE