Achievements
We developed and optimized a 2D biosensor, which was able to detect IPTG and 3-oxo-C12-HSL during artificial induction and we demonstrated the ability to detect living cells of the pathogen Pseudomonas aeruginosa using the same 2D biosensor. To ensure that the fluorescence signal resulted from the sensor constructs build in the sensor chips and not from the medium or E. coli itself [http://parts.igem.org/Part:BBa_B0015 B0015] in NEB was used as negative control during sensor chip induction with IPTG, HSL and P. aeruginosa (Negativ control, displayed below).
Testing our Sensor Chips in a Plate Reader
To establish a prove of principle for our sensor chip design we used our construct [http://parts.igem.org/Part:BBa_K1319042 K1319042], which incorporates an IPTG inducible iLOV. E. coli cells containing the latter construct were introduced into sensor chips and fluorescence was measured every 15 minutes subsequently to induction with 2 µl 100 mM IPTG (Testing K1319042 in our sensor chips, displayed on the left).
We observed a distinct difference in fluorescence between the non-induced chip (top) and the induced chip (bottom). The middle of the bottom chip started to exhibit a clear and visible fluorescence and the fluorescence increased over time and spread outward. The top chip, however, also showed increase in the measured fluorescence over time which was primarily due to a leaky promoter and background fluorescence.
Detecting 3-oxo-C12-HSL with Sensor Chips
In an initially attempt to detect 3-oxo-C12-HSL, we incorporated the [http://parts.igem.org/Part:BBa_K131026 K131026] construct generated by the 2008 iGEM Team Calgary in our sensor chips. The latter construct generates a fluorescent signal based on GFP in presence of 3-oxo-C12-HSL molecules produced by P. aeruginosa during quorum sensing (Jimenez, Koch, Thompson et al., 2012). First, we tested the construct by direct induction of 3-oxo-C12-HSL (0.2 µl, 500 µg/ml) whereby fluorescence measurement was taken every 15 minutes with an excitation wavelength of 496 nm and an emission wavelength of 516 nm (Testing K131026 in our sensor chips, displayed on the rigth).
A distinct fluorescence signal was observed on the induced chip (bottom) compared to the non-induced chip (top).
Fluorescence started in the middle of the chip (point of induction) and then extended outwards, still showing an increasing fluorescence signal. The basal level of fluorescence was attributed to leakiness of the promoter and general background fluorescence of growing E. coli cells. In the induced chip (bottom), the background fluorescence was lower than in the non-induced chip (top), because the signal masked the noise. The difference between the induced and non-induced chips indicated a clear response to the HSL and proofed the ability of our 2D sensor chip design to detect HSLs produced by Pseudomonas aeruginosa.
Detecting IPTG with sensor chips
The video displayed on the left side (Detecting IPTG with sensor chips) shows the construct [http://parts.igem.org/Part:BBa_I746909 I746909] from the 2007 iGEM Team Cambridge, which expresses super folder GFP under the control of a T7 promoter. This construct was introduced into BL21(DE3) cells making the expression IPTG inducible due to the T7 RNA Polymerase encoded in the genome of BL21(DE3) under the control of a lacI promoter.
While the left chip does not show visible fluorescence, the right chip exhibits a strong fluorescence signal. This elucidates the ability of the sensor chip technology to detect IPTG. The fluorescence response is also high enough to be detected and analyzed by our measurement device WatsOn.
Detecting the 3-oxo-C12-HSL with K131026 in our Sensor Chips with WatsOn
The next step towards the final goal to detect Pseudomonas aeruginosa was to replicate the detection of 3-oxo-C12-HSL, which was established in the plate reader, in our own WatsOn device. Therefore, we again used K131026 as our construct in E. coli BL21(DE3) cells and induced with 0.2 µL 3-oxo-C12-HSL with a concentration of 500 µg/mL. The right chip was induced and - as a negative control - the left chip was not induced. Pictures were taken every 4 minutes.
The result was a clear replication of the success of the plate reader experiment. The induced chip shows a clear fluorescence response eminating from the center where the induction with HSL took place. This demonstrates the ability of not only our sensor chips but also our measurement device WatsOn to successfully detect 3-oxo-C12-HSL.
Detecting Pseudomonas aeruginosa with K131026 in our Sensor Chip with WatsOn
After establishing the successful detection of 3-oxo-C12 with our sensor chips the next step was the detection of Pseudomonas aeruginosa with our measurement device WatsOn. Therefore sensor chips with K131026 were again prepared and the right chip was induced with 0.2 µl of a Pseudomonas aeruginosa culture while the left chip was not induced.
The results clearly demonstrate our ability to detect Pseudomonas aeruginosa with our measurement device WatsOn. On the induced chip a definite fluorescence response is visible in response to Pseudomonas aeruginosa. The fluorescence eminates outward from the induction point and shows a significant difference to the non induced chip. Therefore detection of Pseudomonas aeruginosa is possible with our sensor chip technology in our measurement device WatsOn!
|