Team:Aachen/Notebook/Protocols/Analytical methods
From 2014.igem.org
Aschechtel (Talk | contribs) (→Bradford Assay) |
(→Sodium dodecyl sulfate polyacrylamide gel electrophoresis (SDS-PAGE)) |
||
(10 intermediate revisions not shown) | |||
Line 52: | Line 52: | ||
# mix with 1 µl loading dye | # mix with 1 µl loading dye | ||
# apply onto agarose gel together with a marker | # apply onto agarose gel together with a marker | ||
- | # run at | + | # run at 100 V for 35 minutes for a full gel |
- | == Sodium | + | == Sodium Dodecyl Sulfate Polyacrylamide Gel Electrophoresis (SDS-PAGE) == |
- | The SDS-PAGE | + | The SDS-PAGE is used to determine certain features of the cells' proteome such as the strength of expression of a desired protein. |
'''Cell Preparation''' | '''Cell Preparation''' | ||
Line 61: | Line 61: | ||
* centrifuge for 15 min at 13.000 rpm | * centrifuge for 15 min at 13.000 rpm | ||
* mix the supernatant with 2x lammli buffer with β-mercaptoethanol | * mix the supernatant with 2x lammli buffer with β-mercaptoethanol | ||
- | * denatured for 5 min at | + | * denatured for 5 min at 95°C |
* sample to the gel | * sample to the gel | ||
Line 70: | Line 70: | ||
'''1.5x Buffer''' | '''1.5x Buffer''' | ||
* 1.5 M Tris-Cl pH = 8.8 | * 1.5 M Tris-Cl pH = 8.8 | ||
- | * in 1 L is 40 ml 10 | + | * in 1 L is 40 ml 10% SDS |
'''Gels''' | '''Gels''' | ||
Line 76: | Line 76: | ||
{| class="wikitable" style="text-align: right;" | {| class="wikitable" style="text-align: right;" | ||
! | ! | ||
- | !! style="border-left: 2px solid #404040;" colspan="3"|0.75 mm 12 % RUNNING Gel | + | !! style="border-left: 2px solid #404040;" colspan="3"|0.75 mm 12% RUNNING Gel |
- | !! style="border-left: 2px solid #404040; background-color:#8ebae5;" colspan="3"|1 mm 4 % STACKING Gel | + | !! style="border-left: 2px solid #404040; background-color:#8ebae5;" colspan="3"|1 mm 4% STACKING Gel |
|- | |- | ||
| | | | ||
Line 91: | Line 91: | ||
| style="border-left: 2px solid #404040;"| 0.65 ml || 1.3 ml || 2.6 ml | | style="border-left: 2px solid #404040;"| 0.65 ml || 1.3 ml || 2.6 ml | ||
|- | |- | ||
- | | '''30 % Acrylamide (37.5:1)''' | + | | '''30% Acrylamide (37.5:1)''' |
| style="border-left: 2px solid #404040;"| 2 ml || 4 ml || 8 ml | | style="border-left: 2px solid #404040;"| 2 ml || 4 ml || 8 ml | ||
| style="border-left: 2px solid #404040;"| 0.325 ml || 0.65 ml || 1.3 ml | | style="border-left: 2px solid #404040;"| 0.325 ml || 0.65 ml || 1.3 ml | ||
|- | |- | ||
- | | '''10 % APS''' | + | | '''10% APS''' |
| style="border-left: 2px solid #404040;"| 50 µl || 100 µl || 200 µl | | style="border-left: 2px solid #404040;"| 50 µl || 100 µl || 200 µl | ||
| style="border-left: 2px solid #404040;"| 25 µl || 50 µl || 100 µl | | style="border-left: 2px solid #404040;"| 25 µl || 50 µl || 100 µl | ||
Line 121: | Line 121: | ||
== Measurement of Fluorescence == | == Measurement of Fluorescence == | ||
- | The measurement of fluorescence was performed using the Synergy Mx (BioTek) | + | The measurement of fluorescence was performed using the Synergy Mx Microplate Reader (BioTek) and the corresponding Gen5 2.01 software. Unless stated otherwise following parameters were used: |
- | + | * volume of sample in each well: 100 µl | |
- | * volume of sample in each well: | + | * measurement of GFP/eGFP/sfGFP |
- | * | + | ** excitation wavelength: 496 ± 9 nm |
+ | ** emission wavelength: 516 ± 9 nm | ||
+ | * measurement of iLOV | ||
+ | ** excitation wavelength: 450 ± 9 nm | ||
+ | ** emission wavelength: 495 ± 9 nm | ||
+ | The fluorescence is given out in arbitrary units (a.u.). | ||
== Measurement of Optical Density == | == Measurement of Optical Density == | ||
- | Depending on the number of samples, two different devices were used for measurement of | + | Depending on the number of samples, two different devices were used for measurement of OD, the Unico Spectrophotometer 1201 (Fisher Bioblock Scientific) and the Synergy Mx Microplate Reader (BioTek) with the corresponding Gen5 2.01 software. |
+ | |||
+ | Unless stated otherwise following parameters were used for the plate reader measurements: | ||
+ | * volume of sample in each well: 100 µl | ||
+ | * OD<sub>600</sub> with ''pathlength correction'' (including the height of liquid into the calculation of the optical density) | ||
+ | |||
+ | The respective culture medium was used as reference. | ||
Latest revision as of 03:42, 18 October 2014
|