Team:Aachen/Notebook/Protocols/detection
From 2014.igem.org
Aschechtel (Talk | contribs) (→Chip Production) |
(→Measurement of Fluorescence) |
||
(2 intermediate revisions not shown) | |||
Line 53: | Line 53: | ||
# over night culture of sensor cells (50 ml in a 250 ml flask with) max. 16 h | # over night culture of sensor cells (50 ml in a 250 ml flask with) max. 16 h | ||
- | # centrifuge all 50 ml by 3000 g for 10 min at RT ( | + | # centrifuge all 50 ml by 3000 g for 10 min at RT (21°C). |
# discard the supernatant | # discard the supernatant | ||
- | # re-suspend the pellet in 1 mL tempered (~ | + | # re-suspend the pellet in 1 mL tempered (~21°C) LB-medium. |
'''Agar Preparation''' | '''Agar Preparation''' | ||
- | # autoclave 50 ml medium with 1.5 | + | # autoclave 50 ml medium with 1.5%(w/v) agarose (has to be multiplied with the number of chips prepared). |
- | # cool it down to | + | # cool it down to 45°C in a water bath. |
Line 71: | Line 71: | ||
# cut out the chips with a scalpel. | # cut out the chips with a scalpel. | ||
# put two chips into a labeled petri dish and store additional 4 chips in labeled petri dishs in the refrigerator. | # put two chips into a labeled petri dish and store additional 4 chips in labeled petri dishs in the refrigerator. | ||
- | # incubate two chips for 1 h at | + | # incubate two chips for 1 h at 37°C prior to induction. |
Line 84: | Line 84: | ||
'''First''', the '''Gel Doc™ XR+''' (BIO-RAD) was used, exciting with UV light for an exposure time of 1 s. | '''First''', the '''Gel Doc™ XR+''' (BIO-RAD) was used, exciting with UV light for an exposure time of 1 s. | ||
- | '''Second''', we used our own device [https://2014.igem.org/Team:Aachen/Project/Measurement_Device '''WatsOn'''] with blue light for excitation (450 or 480 nm) and special filters infront of the camera for selecting the appropriate emission spectrum. You can read even more about building your own WatsOn [https://2014.igem.org/Team:Aachen/Notebook/Engineering/WatsOn here]. | + | '''Second''', we used our own device [https://2014.igem.org/Team:Aachen/Project/Measurement_Device '''''WatsOn'''''] with blue light for excitation (450 or 480 nm) and special filters infront of the camera for selecting the appropriate emission spectrum. You can read even more about building your own ''WatsOn'' [https://2014.igem.org/Team:Aachen/Notebook/Engineering/WatsOn here]. |
'''Third''', we used the '''Synergy Mx microplate reader''' (BioTek), putting the sensor chips into the lid of a common clear well plate. GFP was measured with an excitation of 496 ± 9 nm and an emission of 516 ± 9 nm and iLOV with an excitation of 450 ± 9 nm and emission of 495 ± 9 nm. | '''Third''', we used the '''Synergy Mx microplate reader''' (BioTek), putting the sensor chips into the lid of a common clear well plate. GFP was measured with an excitation of 496 ± 9 nm and an emission of 516 ± 9 nm and iLOV with an excitation of 450 ± 9 nm and emission of 495 ± 9 nm. |