Team:DTU-Denmark/Interlab form

From 2014.igem.org

(Difference between revisions)
Line 2,621: Line 2,621:
<p>
<p>
Histograms of fluorescence values for each construct are shown below. At least 100,000 cells were measured for each construct. Values on the X-axes are log10-values of the fluorescence measurements.</br>
Histograms of fluorescence values for each construct are shown below. At least 100,000 cells were measured for each construct. Values on the X-axes are log10-values of the fluorescence measurements.</br>
-
For each construct threshold was determined to split the measurements into high-fluorescence and low-fluorescence cells. The high-fluorescence sub-datasets were log10-transformed and a normal distribution was fitted. Mean and standard deviation for the log-normal distributions are reported below each histogram.
+
For each construct, a threshold was determined to split the measurements into high-fluorescence and low-fluorescence cells. The high-fluorescence sub-datasets were log10-transformed and a normal distribution was fitted. Mean and standard deviation for the log-normal distributions are reported below each histogram. The vertical red line indicates the threshold.
</p>
</p>

Revision as of 21:05, 23 September 2014

iGEM 2014 Measurement Interlab Study Worksheet

Inter-Lab Study Worksheet:

For each assay that you perform, fill out the following worksheet. Answer each question with enough detail to allow another person to replicate your measurements without needing to ask you any questions. This does not necessarily mean you need to describe everything in detail---for example, if you use a standard assay, you just need to give enough information to allow another person to use that assay in the same way that you did.

 

Section I: Provenance & Release

Who did the actual work to acquire these measurements?

Our laboratory group carried out the experiments this group consists of Anne Sofie Lærke Hansen, Anne Pihl Bali, Caroline Mosbech, Kristian Barreth Karlsen, Kristian Jensen & Rasmus Bech

 

What other people should be credited for these measurements?

We did all the experimental work for the interlab study ourselves. We did

Ali Altintas assisted us in setting up the BioLector. Anne Egholm helped us to use the Flow Cytometer.

 

On what dates were the protocols run and the measurements taken?

The applied BioLector experiment ran: 10th-11th of September 2014

The applied Flow cytometry experiment was done on the 22nd of September 2014.

 

Do all persons involved consent to the inclusion of this data in publications derived from the iGEM interlab study?

Yes.


 

Section II: Protocols

What protocol did you use to prepare samples for measurement?

Construct strains:

Create multiple strains expressing GFP from different promoters from the Andersen Library. Constructs were prepared with Standard Assembly. (see protocol).

Plasmid were purified with Zyppy™ Plasmid Miniprep Kit from Zymo research following manufacturer's protocol.

Competent DH5&alpha; cells were prepared (see protocol).

Purified plasmids were transformed into competent DH5&alpha; cells (see protocol).

 

Culture measurements:

We used cell culture tubes (10 mL) containing 2 mL LB medium (with 0.5% NaCl) and 25 &mu;g/mL chloramphenicol for each strain. To each of the tubes, we inoculated cells from a single colony with a sterile inoculation loop and left it overnight at 37&deg;C, shaking at 350 rpm.

The wells of the sterile BioLector plate (48 well FlowerPlate) was prepared with 1450 &mu;L (1500&mu;L for controls) of LB (0.5% NaCl) containing 25 &mu;g/mL chloramphenicol. The following day we transferred 10 &mu;L of each culture into an Eppendorf tube with 990 &mu;L of sterile physiological salt water (creating a 100-fold dilution). We inoculated the individual BioLector wells with 10 &mu;L of the dilution (creating an additional 150-fold dilution). Each strain was grown in triplicates to make sure we had 3 independent measurements and a more precise result. We incubated the plate in the BioLector at 37&deg;C and 1000 rpm for 12.5 hours (stationary phase was reached). During the growth experiment we measured biomass and GFP fluorescence every 5 minutes. Biomass was measured by light scatter at 620 nm, while fluorescence was measured at 520 nm, with an excitation wavelength of 488 nm.

 

Single cell measurements:

2 mL LB (0.5% NaCl) was inoculated with cells from a plate and grown overnight.

The cultures were washed and resuspended in PBS.

The cell suspensions were diluted in PBS to a concentration where the flow cytometer could detect 1000-1500 events per second.

 


 

What sort of instrument did you use to acquire measurements?

What is the model and manufacturer?

Instrument: BioLector (Basic) - Microbioreactor System; Manufacturer: m2p-laps.

Instrument: FACSCalibur Cell Analyzer; Manufacturer: BD Biosciences

 

How is it configured for your measurements?

The BioLector is fitted with filters for measuring biomass (scatter at 620 nm) and GFP (emission at 520 nm, excitation with 488 nm). Settings: Gain: GFP: 80; Biomass: 30; Humidity: controlled at 95%; Shake: 1000 rpm.

The FACSCalibur was used with settings…(Jensen thinks about it… 4.5 du kan husked et XXXXX)

 

What protocol did you use to take measurements?

The BioLector was instructed to take measurements every 4.86 minutes. Settings: GFP: 80; Biomass: 30; Humidity: controlled at 95%. 2.5 mL LB medie with proper antibiotics in each well.

 

The FACS was set up with a threshold on side scatter, as events with low side-scatter are very likely noise, i.e. not actual cells. Each sample was then analysed at 1000-1500 events per second, until a total of 100,000 events was reached.

 

What method is used to determine whether to include or exclude each sample from the data set?

Strains that showed correct digest patterns were included. Strains that were sequenced and showed wrong sequences were excluded. As were cultures that did not grow, and single replicates that showed negligible fluorescence.

 

What exactly were the controls that you used?

DH5&alpha; with BBa_E0240 (no promoter) was used to determine cell background fluorescence (without GFP). For the BioLector, clean media (without cells) was used to determine media fluorescence.

 

What quantities were measured?

On the BioLector: GFP fluorescence (488 nm /520 nm) and biomass (scatter at 620nm). The flow cytometer measured forward scatter (cell size), side scatter (cell granularity) and fluorescence.

 

How much time did it take to acquire each set of measurements?

The running time of the BioLector experiment was 12 hours and 30 minutes. One cycle took 4.86 min.

Flow cytometry: The measurements themselves took less than 1 hour.

 

How much does it cost to acquire a set of measurements?

Assuming the lab already have a BioLector. For a full plate with samples: Flowerplate for the BioLector, 170mL LB medium, chloramphenicol, 16 pts. inoculation tubes, tips, Eppendorf tubes approximately $150.

 

What are the practical limits on the number or rate of measurements taken with this instrument and protocol?

BioLector: With the used protocol 48 samples can be measured within short intervals of time, however, 3 of these should be used as control. So when measuring triplicates it is possible to measure 15 different strains in the protocol used for the interlab study. Analysing such a plate takes approximately 12 hours.

 

Section III: Measured Quantities

For each type of quantity measured, report on the following:

What are the units of the measurement?

All measurements are in arbitrary units in both the BioLector and the FACSCalibur.

 

What is the equivalent unit expressed as a combination of the seven SI base units?

Not applicable due to arbitrary units.

 


 

Precision:

What is the range of possible measured values for this quantity, using your instrument as configured for these measurements?

For the BioLector the linear range for OD 600 is between 0.2 and 600.

Linear range for GFP XXXXX

 

What are the significant figures for these measurements?

We have estimated a mixed model with promoter as fixed effect and replicates as random effects, this model showed the residual standard deviation as 0.040. With this result we can say that a single measurement is no more precise than the measured value ± 0.040.

 

Is the precision the same across the entire range? If not, how does it differ?

We have no data to determine if the precision differ through the entire range.

 

How did you determine these answers?

Jensen thinks about something

 

Accuracy:

When was the instrument last calibrated?

The machine “self-calibrate” before every cycle, meaning the BioLector sensor/probe returns to the original position and measure a fixed known light intensity. It is approximately 5 years old but the filters have been changed 3 years ago.

 

How was the instrument calibrated?

The machine “self-calibrate” before every cycle, meaning the BioLector sensor/probe returns to the original position and measure a fixed known light intensity.


 

Section IV: Measurements


BioLector experiment


For each sample, report:

The following table shows the BioMass and Fluorescence measurements acquired from the BioLector. Each entry in the table is the mean value of all the measurements taken during a 3-hour period, with 4.88 min intervals. The full dataset can be found here XX.

Name

Biomass

GFP fluorescence

Fluorescence/OD

Note

J23100-1

12.59

2.13

0.1695

Excluded

J23101-1

54.80

30.17

0.5506

J23102-1

33.43

1.27

0.0381

Excluded

J23103-1

62.60

0.38

0.0061

J23106-1

59.10

30.66

0.5188

J23107-1

60.78

13.95

0.2296

J23100-2

57.97

47.45

0.8185

J23101-2

59.11

31.42

0.5315

J23102-2

61.60

18.93

0.3073

J23103-2

65.37

0.42

0.0064

J23106-2

59.04

30.29

0.5130

J23107-2

60.63

13.82

0.2279

J23100-3

61.51

49.65

0.8072

J23101-3

60.83

32.54

0.5349

J23102-3

61.59

21.08

0.3422

J23103-3

65.75

0.39

0.0060

J23106-3

59.63

29.42

0.4934

J23107-3

60.89

13.89

0.2280

J23113-1

69.70

0.43

0.0062

J23114-1

69.60

1.56

0.0225

J23115-1

66.27

1.85

0.0279

J23118-1

50.75

95.47

1.8811

J23119-1

46.99

17.50

0.3724

No promoter-1

64.81

0.39

0.0060

I20260-1

61.36

19.75

0.3219

J23113-2

72.55

0.45

0.0062

J23114-2

71.19

1.58

0.0223

J23115-2

68.07

1.88

0.0276

J23118-2

50.15

92.32

1.8410

J23119-2

47.35

17.18

0.3629

No promoter-2

67.06

0.38

0.0057

I20260-2

62.79

18.62

0.2966

J23113-3

77.21

0.43

0.0056

J23114-3

76.52

1.63

0.0213

J23115-3

72.29

1.92

0.0265

J23118-3

52.00

96.50

1.8556

J23119-3

47.13

16.81

0.3567

No promoter-3

70.07

0.39

0.0056

I20260-3

64.47

17.33

0.2688

 

For each group of replicates, report:

The data was analyzed by fitting a statistical mixed model to the data, using the lmerTest package in R. The different promoters were modeled as a fixed effect, and replicates were modeled as a random effect. The estimated means for each promoter and the standard error of the mean can be seen in the table and chart below.

Strain

Mean

Standard Error

Background-subtracted mean

J23100

0.91

0.01300

0.9050

J23101

0.61

0.01160

0.5980

J23102

0.37

0.01300

0.3610

J23103

0.01

0.01160

0.0005

J23106

0.58

0.01160

0.5690

J23107

0.26

0.01160

0.2500

J23113

0.01

0.01160

0.0003

J23114

0.03

0.01160

0.0183

J23115

0.03

0.01160

0.0243

J23118

2.14

0.01160

2.1300

J23119

0.42

0.01160

0.4120

I20260

0.32

0.01160

0.3080

E0240

0.0079

0.00821

The background-subtracted values were calculated by subtracting the mean of E0240 from the other means.

The replicate standard deviation was estimated to be 0.014, meaning that the effect of using multiple replicates contributes a random effect distributed with this standard deviation.
The residual standard deviation was estimated to be 0.040, meaning that the random effect of taking multiple measurements on the same sample has this standard deviation.


Which, if any, of the samples are excluded and why

In two of the triplicate sets, one outlier was excluded: For the promoter J23100, one replicate did not grow very much. For promoter J23102 one replicate had negligible fluorescence.


Flow cytometry

Histograms of fluorescence values for each construct are shown below. At least 100,000 cells were measured for each construct. Values on the X-axes are log10-values of the fluorescence measurements.
For each construct, a threshold was determined to split the measurements into high-fluorescence and low-fluorescence cells. The high-fluorescence sub-datasets were log10-transformed and a normal distribution was fitted. Mean and standard deviation for the log-normal distributions are reported below each histogram. The vertical red line indicates the threshold.

Mean: 2.41
SD: 0.25


Mean: 1.02
SD: 0.27


Mean: 3.05 SD: 0.24


Mean: 2.32 SD: 0.48


Mean: 2.33 SD: 0.20